Electrical Shock |
Type BF Applied Part. |
Environmental Conditions |
Transport and store 50°F to 131°F (10°C to 55°C). Operate 41°F to 104°F (5°C to 40°C). Humidity less than 90%. Atmospheric pressure from sea level to 9,842 feet (3,000m). |
Ingress of Water |
Not protected against ingress of water. |
Flammability |
Do not use around flammable gases, liquids or materials. |
Mode of Operation |
Continuous. |
Dimensions |
6.1” x 3.5” x 1.9” (15.5 x 8.9 x 4.8cm) |
Weight |
.4 lbs. (.18kg) |
Cleaning |
Clean the case and lead clip wires as needed with an alcohol moistened cloth. Do not immerse in fluids. |
Disposal |
Dispose of according to local, state and federal regulations. Remove battery before disposal. |
Controls |
Two (dose and current). |
Dose Range |
0 to 80 mA-min. |
Maximum Voltage |
80V DC |
Maximum Current |
4.0 mA |
Current Ramp Up |
Automatic (0 to 4.0 mA). Built-in option for manual override. |
Current Ramp Down |
Automatic at end of treatment; paused or turned off by depressing knob. Built-in option for manual override |
Battery |
Use only 9V DC Alkaline. Ensure battery door is in place before starting treatment. Remove battery from unit when not in use. |
Display |
Dose, Time Remaining, and Current (displayed simultaneously with interactive set-up). |
Pause Feature |
YES (w/recalculation when restarted) |
Visual Indicators |
Low battery and open circuit. |
Audible Alerts |
Low battery, open circuit, and end-of treatment. |
Auto Shut-Off |
Will automatically shut off after one (1) minute, if not in use. |
activadose®II