setstats

activadose

activadose
Electrical Shock
Internal powered Type BF Equipment.
Environmental Conditions
Transport and store in a cool dry place between
68°F to 122°F (20°C to 50°C), less than 90%, humidity and operational up to 32,808’ (10,000m) above sea level.
Ingress of Water
Not protected against ingress of water.
Flammability
Do not use around flammable gases, liquids or materials.
Mode of Operation
Continuous.
Dimensions
6.52” x 3.5” x 2.25” (16 x 8.9 x 5.7cm)
Weight
.53 lbs. (.24kg)
Cleaning
Clean the case and lead clip wires as needed with an alcohol moistened cloth. Do not immerse in fluids.
Disposal
Dispose of according to local, state and federal regulations. Remove battery before disposal.
Controls
Two (dose and current).
Dose Range
0 to 80 mA-min.
Maximum Voltage
80V DC
Maximum Current
4.0 mA
Current Ramp Up
Automatic (0 to 4.0 mA). Built-in option for manual override.
Current Ramp Down
Automatic at end of treatment. Built-in option for manual override.
Battery
Use only 9V DC Alkaline. Ensure battery door is in place before starting treatment. Remove battery from unit when not in use.
Display
Dose, Time Remaining, and Current (displayed simultaneously with interactive set-up).
Pause Feature
YES (w/ reset to zero).
Visual Indicators
Low battery and open circuit.
Audible Alerts
Low battery, open circuit, and end-of treatment.